Model : SEHY-CT750
Features :
- Measurement display mode: positive wave, negative wave,full wave and wave ;
- It has a linear suppression function, the maximum inhibition is 80% of the screen height;
- It can switch between normal probe, dual-crystal probe and through probe;
- It has a gate setting and alarm function. The position and width of the gate can be arbitrarily set on the screen, and an incoming wave alarm can be set;
- 500 independent detection channels, each channel set a separate set of testing parameters, DAC curve;
- Two display modes with angle and K value;
- The DAC curve is generated automatically, and up to 10 points can be recorded, with four additional adjustable offset curves;
- AVG curve is generated automatically, and two types of defects can be customized;
- Automatic probe calibration function;
- With storage function, it can store 100 A sweep patterns, parameters and DAC curve;
- Has a storage graphics playback function, the stored A-scan graphics from the storage area and displayed on the screen;
- Has a delete function to delete the specified content (represented by the storage group number) from the storage area;
- Peak memory function;
- Freeze and thaw functions with waveform and inspection parameters;
- With sound path measurement and echo frequency analysis function;
- Real-time power status indication function;
- Support wireline communication interface;
- Support USB disk to transfer data, easy to save the data to PC
- Li battery, which can work more than ten hours
Specifications :
Specifications | Technical data |
Scan range (mm) | Scan range (mm):0~10000 |
Grades: 2.5,5,10,20, 30,40,50,60,70,80,90, 100,150,200, 250, 300, 350, 400, 450,500,600,700,800,900,1000,2000,3000,4000,5000,6000,7000,8000,9000, 10000. | |
Adjusting step distance: 1mm | |
Pulse shift ( ms ) | Pulse shift ( ms ): -7 to +9984 |
Grades: -20, -10, 0.0, 10, 20, 50, 100, 150, 200, 250, 300, 350, 400, 450, 500, 600,700,800,900,1000,1500,2000,2500,3000,3400 | |
700,800,900,100,015,000,000,000,000,000,000.00 | |
Adjustment step: 1 (-7 ms to 9984 ms) | |
Probe zero point(ms) | Probe zero point:0.0~99.99 |
Adjust the step:0.01 | |
Material sound speed(m/s) | Material sound speed:1000~15000 |
7 fixed sound speeds:2260,2730,3080,3230,4700,5900,6300 | |
Adjust the step:1 | |
Working methods | Single probe (receive and send), dual probes (one for receiving one for sending) |
Frequency Range(MHz) | 0.5~20 |
Gain adjustment(dB) | 0~120 |
Adjust the step:0.2,1,2,4,6,8,10,20,50 | |
Linear inhibition | Screen height 0% to 80%, step size: 1% |
Vertical linearity error | Vertical linear error is no more than 3% |
Horizontal linear error | Within the scan range, no more than 0.2% |
Flaw detection sensitivity margin | ³60dB |
Dynamic Range | ³32dB |
Alarm | Incoming wave alarm |
Display screen | Display:High-brightness color flat panel display |
A-Scan display area | Full screen or local |
A-Scan shows freeze and thaw A-Scan fills | |
Flaw detection channel | 500 |
data storage | 100 A-Scan graphics |
PC communication interface standard | USB |
Units of measurement | Mm |
Power Adapter | Input 100V~240V/50Hz~60Hz |
Output 13V/5A | |
Battery | Lithium (Li) Battery 5000mAh |
Working temperature (°C) | -10~50°C |
Working humidity (RH) | 20%~90% |
Interface Type | BNC |
Dimensions (mm) | 246*155*55 |
Weight(kg) | 1.3 |
Packing List :
Inquiry - Ultrasonic Flaw Detector SEHY-CT750